Chapter 2 167
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
Example: MCP:OFFS:LIST:BWIDth 5MHz,3MHz,5MHz
MCP:OFFS:LIST:BWID?
2.16.4.4 Upper Offset Limit
Sets the limit for the selected offset (refer to “Offset” above) to the right of the carriers.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in instrument state.
Factory Preset: 0 dB, 0 dB, 0 dB
Default Terminator: dB
Knob Increment: 1
Step Key
Increment: 5
Range: –200 dB to 200 dB
History: Added with PSA firmware revision A.03.00
Added with ESA firmware revision A.08.00
Remote Command:
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA <dB>,<dB>,<dB>
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA?
Example: CALC:MCP:OFFS:LIST:LIM:POS:DATA 50dB,55dB,60dB
:CALC:MCP:OFFS:LIST:LIM:POS:DATA?
2.16.4.5 Lower Offset Limit
Sets the limit for the selected offset (refer to “Offset” above) to the left of the carriers.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in instrument state.
Factory Preset: 0 dB, 0 dB, 0 dB
Default Terminator: dB
Knob Increment: 1
Step Key
Increment: 5
Range: –200 dB through 200 dB
History: Added with PSA firmware revision A.03.00