34 Chapter 2
One-Button Measurement Functions
Mode Setup (Spectrum Analysis Mode)
One-Button Measurement Functions
2.1.1.1 None
Selects no radio standard. If Radio Std, None is selected when a measurement is running,
all measurement parameters are returned to their default values. Analyzer parameters
outside the MEASURE or Meas Setup key menus are not affected.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
History: Modified with PSA firmware revision A.04.00
Added with PSA firmware revision A.02.00
Modified with ESA firmware revision A.08.00
Added with ESA firmware revision A.07.00
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page33.
Example: RAD:STAN NONE
2.1.1.2 IS-95A
Sets the specific parameters for the selected measurement (located under the “MEASURE
(Spectrum Analysis Mode)” key description) appropriate for industry standard IS-95A. The
Spectrum Emissions Mask, Harmonic Distortion, and Spurious Emissions measurements
are not available when this radio standard is selected.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
History: Modified with PSA firmware revision A.04.00
Added with PSA firmware revision A.02.00
Modified with ESA firmware revision A.08.00
Added with ESA firmware revision A.07.00
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page33.
Example: RAD:STAN IS95
2.1.1.3 J-STD-008
Sets the specific parameters for the selected measurement (located under the “MEASURE
(Spectrum Analysis Mode)” key description) appropriate for industry standard J-STD-008.
The Spectrum Emissions Mask, Harmonic Distortion, MCP, and Spurious Emissions
measurements are not available when this radio standard is selected.