166 Chapter 2
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
Factory Preset: 5 MHz, 10 MHz, 15 MHz
Terminators: Hz, kHz, MHz, GHz
Default Terminator: Hz
Range: 0 MHz to 45 MHz
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.09.00
Remote Command:
[:SENSe]:MCPower:OFFSet:LIST:[FREQuency]<Hz>,<Hz>,<Hz>
[:SENSe]:MCPower:OFFSet:LIST:[FREQuency]?
Example: MCP:OFFS:LIST:5MHz,7.5MHz,15MHz
MCP:OFFS:LIST:?
2.16.4.3 Offset Integ BW
Sets the bandwidth over which the power is integrated for the selected offset (refer to
“Offset” above).
Key Path: Meas Setup, Offset/Limits
Key Notes: If RRC Filter is set to On, the actual integration bandwidth used is the
displayed integration bandwidth multiplied by (1 + Filter Alpha).
State Saved: Saved in instrument state.
Factory Preset: 3.84 MHz, 3.84 MHz, 3.84 MHz
Terminators: Hz, kHz, MHz, GHz
Default Terminator: Hz
Knob Increment: Span ÷ 50
Step Key
Increment: If CF Step(Auto) is selected: span/10
If CF Step(Man) is selected: CF Step
Range: 100 Hz to 20 MHz
History: Added with PSA firmware revision A.03.00
Added with ESA firmware revision A.09.00
Remote Command:
[:SENSe]:MCPower:OFFSet:LIST:BANDwidth|BWIDth:[INTegration]<Hz>,<Hz>,<Hz>
[:SENSe]:MCPower:OFFSet:LIST:BANDwidth|BWIDth:[INTegration]<Hz>,<Hz>,<Hz>