Chapter 2 141
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
2.13 Meas Setup (Harmonic Distortion)
When the harmonic distortion measurement has been selected in the Measure key menu of
the Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
Key Path: Front-Panel key
Dependencies/
Couplings: This measurement is not available when external mixing has been
selected.
Saved State: Not saved.

2.13.1 Avg Number

Press Avg Number (On) to specify the number of measurement averages used when
calculating the measurement result. The average will be displayed at the end of each
sweep.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10 averages / Off
Range: 1 to 1000
Remote Command:
[:SENSe]:HARMonics:AVERage:COUNt <integer>
[:SENSe]:HARMonics:AVERage:COUNt?
[:SENSe]:HARMonics:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HARMonics:AVERage[:STATe]?
Example: HARM:AVER:COUN 100
HARM:AVER ON

2.13.2 Avg Mode

Press Avg Mode to select the type of termination control used for the averaging function to
either Exp or Repeat. This determines the averaging action after the specified number of
measurements (average count) is reached.
EXP (Exponential Averaging mode)—When you set Avg Mode to Exp, each successive
data acquisition after the average count is reached is exponentially weighted and