Chapter 2 161
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
carrier power present values.
State Saved: Saved in instrument state.
Factory Preset: Auto/Calculated based on current center frequency. Refer to the algorithm
above when defining the Ref Carrier Freq(Auto) key.
Terminators: Hz, kHz, MHz, GHz
Default Terminator GHz
Resolution: 1 Hz
Knob Increment: Span ÷ 50
Step Key
Increment: If CF Step(Auto) is selected: span/10.
If CF Step(Man) is selected: CF Step
Range: Analyzer minimum to analyzer maximum
History Added with ESA firmware revision A.09.00.
Added with PSA firmware revision A.03.00.
Remote Command:
[:SENSe]:MCPower:RCFRequency<Freq>
[:SENSe]:MCPower:RCFRequency?
[:SENSe]:MCPower:RCFRequency:AUTO OFF|ON|0|1
[:SENSe]:MCPower:RCFRequency:AUTO?
Example: MCP: RCFR 2 GHz
MCP:RCFR?
2.16.3.4 Configure Carriers
Accesses the Config Carriers menu that allows further definition of each carrier.
Key Path: Meas Setup, Carrier Setup
History: Added with ESA firmware revision A.09.00.
Added with PSA firmware revision A.03.00.
Remote Command:
There is no equivalent remote command.
2.16.3.4.1 Carrier
Selects the carrier number you wish to configure.