Chapter 2 155
One-Button Measurement Functions
Meas Setup (Intermod (TOI))
One-Button Measurement Functions
Range: Non-Zero Span of Analyzer
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00
Remote Command:
[:SENSe]:TOIN:FREQuency:SPAN <number>
[:SENSe]:TOIN:FREQuency:SPAN?
Example: TOIN:FREQ:SPAN 20 MHz
TOIN:FREQ:SPAN?
2.15.4 Max Mixer Lvl
Enables you to set the relationship between the highest signal that can be displayed (the
reference level) and the input attenuation. The relationship applies whenever the
Attenuation is set to Auto. The relationship is that the attenuation is given by reference
level minus the max mixer level. For example, as the reference level changes, the input
attenuator changes to ensure that a signal at the reference level does not exceed the Max
Mixer Lvl setting.
When the TOI measurement is on, this key controls the maximum mixer level,
independent of the previous setting of Max Mixer Lvl located under Amplitude. When the
TOI measurement is Off, the previous maximum Max Mixer Lvl is restored.
Setting Max Mixer Lvl to Auto sets the maximum mixer level to 30 dBm.
Key Path: Meas Setup
Factory Preset: –30.00 dBm/Auto
Range: 0 VDC, +30 dBm
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00
Remote Command:
[:SENSe]:TOIN:FREQuency:MIXer:RANGe[:UPPer] <number>
[:SENSe]:TOIN:FREQuency:MIXer:RANGe[:UPPer]?
[:SENSe]:TOIN:FREQuency:MIXer:RANGe:AUTO OFF|ON|0|1
[:SENSe]:TOIN:FREQuency:MIXer:RANGe:AUTO?
Example: TOIN:FREQ:MIX:RANG –10 dBm
TOIN:FREQ:MIX:RANG?