118 Chapter 2
One-Button Measurement Functions
Meas Setup (Complementary CumulativeDistribution Function—CCDF)
One-Button Measurement Functions
Remote Command Notes: The entered value is rounded to the nearest Res BW filter
frequency.
This command is coupled with the analyzer resolution bandwidth
Example: PST:BAND 10 or PST:BWID 10
PST:BAND? or PST:BWID?
2.7.2 Counts
Press Counts to set the accumulated number of sampling points for data acquisition. The
range is 1 kpoints (kilopoint–1000 points) to 2.000000 Gpoints (Giga point–109 point) wi th
1 kpoint resolution. While this key is activated, enter a value from the numeric keypad by
terminating with one of the unit keys shown.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 1.00e5
Range: 1e3 to 1e9
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.07.00.
Remote Command:
[:SENSe]:PSTatistic:COUNts <integer>
[:SENSe]:PSTatistic:COUNts?
Example: PST: COUN 1.0e6
PST:COUNt?
2.7.3 Meas Interval
Press Meas Interval to specify the time interval over which the measurement is made. The
minimum setting is 100 µs. The maximum setting depends on the measurement
bandwidth setting but never exceeds 600 ms. The resolution is 1 µs.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 1.0ms
Default Terminator: s
Range: PSA Only: 100 µs to 600 ms