144 Chapter 2
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
2.13.6 Range Table
Enables you to enter the settings for up to 10 ranges, either using the instrument front
panel keys or remotely. Upon entering the range table, the measurement stops, then the
analyzer is set to a constantly sweeping state. The analyzer is then set to the current
values for range 1 (whether range 1 is on or off). If a range is currently off, the values in
the range table for that range are replaced with --- (see Range 5 in the illustration) to
indicate this range is currently inactive.
NOTE You can edit the range table at anytime by pressing any front-panel key
except the following: Esc, System, File, Save, Print Setup, Print, Marker, Peak
Search, Freq Count, Next Marker, Next Window, or Zoom.
Key Path: Meas Setup
Saved State: All values for all ranges are saved in instrument state.
Factory Preset: For each range, the following settings are selected: Range State(On), Zero
Span, Res BW Mode(Man), Video BW(Man) is set to 2 kHz, and Sweep Time
Mode(Auto) is set to 10 ms. The table below defines the other default
parameter settings for the measurement.