134 Chapter 2
One-Button Measurement Functions
Meas Setup (Channel Power—CHP)
One-Button Measurement Functions
Example: CHP:AVER:COUN 10
CHP:AVER:COUN?
CHP:AVER OFF
CHP:AVER?
2.11.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function to
either Exp or Repeat. This determines the averaging action after the specified number of
measurements (average count) is reached.
EXP (Exponential Averaging mode)—When you set Avg Mode to Exp, each successive
data acquisition after the average count is reached is exponentially weighted and
combined with the existing average. Exponential averaging weights new data more
than old data, which facilitates tracking of slow-changing signals. The average will be
displayed at the end of each sweep.
Repeat—When you set Avg Mode to Repeat, after reaching the average count, all
previous result data is cleared and the average count is set back to 1.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: EXPonential
Remote Command:
[:SENSe]:CHPower:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:CHPower:AVERage:TCONrol?
Example: CHP:AVG:TCON EXP
CHP:AVG:TCON?
2.11.3 Integ BW
Press Integ BW to specify the range of integration used in calculating the power in the
channel. Note that the integration bandwidth is displayed on the trace as two markers
connected by an arrow. Be sure the Span of the instrument is set between 1 and 10 times
the integration bandwidth.
Key Path: Meas Setup
Key Notes: If RRC Filter is on, the actual integration bandwidth used is the displayed
integration bandwidth multiplied by (1 + “Filter Alpha”).
State Saved: Saved in instrument state.