214 Chapter 2
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
count) is reached.
EXP (Exponential Averaging mode)—When you set Avg Mode to Exp, each successive
data acquisition after the average count is reached is exponentially weighted and
combined with the existing average. Exponential averaging weights new data more
than old data, which facilitates tracking of slow-changing signals. The average will be
displayed at the end of each sweep.
Repeat—When you set Avg Mode to Repeat, after reaching the average count, all
previous result data is cleared and the average count is set back to 1.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: EXPonential
Range: EXPonential|REPeat
Remote Command:
[:SENSe]:SPURious:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:SPURious:AVERage:TCONrol?
Example: SPUR:AVG:TCON EXP
SPUR:AVG:TCON?
2.24.3 Range Table
Enables you to enter the settings for up to 20 ranges, either using the instrument front
panel keys or remotely. Upon entering the range table, the measurement stops, then the
analyzer is set to a constantly sweeping idle state. The analyzer is then set to the current
values for range 1 (whether range 1 is on or off). If a range is currently off, the values in
the range table for that range are replaced with --- (see Range 5 in the illustration) to