Chapter 2 49
One-Button Measurement Functions
Mode Setup (Spectrum Analysis Mode)
One-Button Measurement Functions
2.1.4 Enable All Measurements
Allows you to access measurements that are disabled. Not all measurements are available
for a particular radio standard. When you select a radio standard, the keys are grayed out
for the measurements that are not available. These disabled measurements do not have
default settings for the standard that you have selected. So you have to configure the
settings yourself when you use them. See also the Retain Params key function.
If you set Enable All Measurements to Yes and you enter one of the measurements that was
disabled, then the Enable All Measurements function will not be available (grayed out). You
cannot turn it off until you enter a measurement that is valid for your selected standard.
Key Path: Mode Setup
Factory Preset: No
History: Added with PSA firmware revision A.04.00
Added with ESA firmware revision A.10.00
Remote Command:
[:SENSe]:RADio:STANdard:EAMeas YES|NO
[:SENSe]:RADio:STANdard:EAMeas?
Example: RAD:STAN:EAM YES
2.1.5 Autorange of Power Setting (Remote command only)
This command is the remote version of the Optimize Ref Level key. If this function is
applicable to the currently selected measurement, this command sets the reference level
and attenuator to optimum values based on the signal present at the input. Channel
Power, Occupied Bandwidth, ACP, Multi Carrier Power, CCDF, Harmonic Distortion,
Burst Power, Intermod (TOI) and SEM measurements are the only supported
measurements for the reference level optimization.
NOTE There is always a minimum of 5 dB for ESA or 6 dB for PSA of attenuation
set to protect the analyzer input.
History: Modified with PSA firmware revision A.04.00
Added with PSA firmware revision A.02.00
Modified with ESA firmware revision A.08.00
Added with ESA firmware revision A.07.00
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example: POW:RANG:AUTO ONCE