216 Chapter 2
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
Remote Command:
There is no equivalent remote command.
2.24.3.1 Range
Selects a range and updates the values on the other Range Table keys so that they reflect
the settings for the selected range. If Range is set to On it is used as part of the
measurement; when set to Off it is excluded.
Key Path: Meas Setup, Range Table
Factory Preset: See “Range Table” on page 214.
Range: Off|On
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted. For example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe OFF|ON|0|1
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe?
Example: SPUR:STAT OFF,ON,ON,OFF,ON,ON
lists values 1 through 6
2.24.3.2 Start Freq
Used to set the start frequency of the analyzer.
Key Path: Meas Setup, Range Table
Factory Preset: See “Range Table” on page 214.
Knob Increment: 1%
Step Key
3 2.100 2.102 100
4 2.175 2.180 100
5 0.800 1.000 4000
6-20 1.5 2.5 Auto
Table 2-6 Range Table Default Settings
Range Start Freq (GHz) Stop Freq (GHz) Res BW (kHz)