164 Chapter 2
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
[:SENSe]:MCPower:CARRier:LIST:WIDTh?
Remote Command Notes:
The position number of the each carrier width in the list sent
corresponds to the carrier number you are defining. For example: “5
MHz, 10 MHz, 5 MHz” defines six carriers. Carriers 1, 2, 4, and 5 are
defined as having power present whereas carriers 3 and 6 do not have
power present. If you need to change carrier 5, you must send all
carriers up to 5. Carrier 6 will remain unchanged. If you send more
values than the number of carriers specified using the Carriers key, (in
this example, there are six carriers) they will be ignored.
The query for this parameter returns the current value for all carriers
(with and without power present).
Example: MCP:CARR:LIST:WIDT 5 MHz,10 MHz,4 MHz,15 MHz,5 MHz,10 MHz
MCP:CARR:LIST:WIDT?
2.16.3.4.4 Carrier IntegBW
Press Carrier IntegBW to define the integration bandwidth used to calculate the power in
the carriers.
Key Path: Meas Setup, Carrier Setup, Configure Carriers
State Saved: Saved in instrument state.
SCPI Status Bits/
OPC Dependencies: The integration bandwidth is specified differently depending on the
radio standard selected. If Mode Setup, Radio Std, 3GPP W-CDMA is selected
and the RRC filter is on, the actual integration bandwidth used will be the
displayed integration bandwidth multiplied by (1 + filter alpha).
Factory Preset: 3.84 MHz, 3.84 MHz, 3.84 MHz, 3.84 MHz
Terminators: Hz, kHz, MHz, GHz
Default Terminator Hz
Resolution:
Knob Increment: Span ÷ 50
Step Key
Increment: If CF Step(Auto) is selected: span/10.
If CF Step(Man) is selected: CF Step
Range: 100 Hz to 500 MHz
History: Added with ESA firmware revision A.09.00.Added with PSA firmware
revision A.03.00.