228 Chapter 2
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
State Saved: Saved in instrument state.
History: Added with PSA firmware revision A.06.00
Remote Command:
See “Detector 2” on page 224.
2.24.3.14.4 MIL Peak Selects peak detection using the MIL specifications.
Key Path: Meas Setup, Range Table, Detector 1 or Detector 2
State Saved: Saved in instrument state.
History: Added with PSA firmware revision A.06.00
Remote Command:
See “Detector 2” on page 224.
2.24.3.14.5 Off Detector 2 is off, there is no second trace.
Key Path: Meas Setup, Range Table, Detector 2
State Saved: Saved in instrument state.
History: Added with PSA firmware revision A.06.00
Remote Command:
See “Detector 2” on page 224.
2.24.3.15 Points
PSA Only: Used to set the points mode of the analyzer.Setting to Man allows you to
determine the setting, or by selecting Auto the analyzer will determine the number of trace
points to ensure the bucket resolution equals RBW/2.
Key Path: Meas Setup, Range Table
State Saved: Saved in instrument state.
Factory Preset: Auto, the value is determined by the radio standard and the instrument
model.
Knob Increment: 1
Step Key
Increment: 1
Range: 101to 8192
History: Added with PSA firmware revision A.06.00