Chapter 2 91
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
Increment: RBW if CF Step is set to Auto with zero span.
Span/10 if CF Step is set to Auto with non-zero span.
CF Step if CF Step is set to Manual.
Range: 300 Hz to 500 MHz. Depends on the adjacent-channel settings and the
minimum and maximum analyzer span.
History: Added with PSA firmware revision A.02.00.
Remote Command:
[:SENSe]:ACPower:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:ACPower:BANDwidth|BWIDth:INTegration?
Example: ACP:BWID:INT 5E6
ACP:BWID:INT?
2.3.4 Offset/Limits
Accesses menu keys that allow you to configure the offsets for the ACP measurement.
Key Path: Meas Setup
2.3.4.1 Offset
Selects the offset the menu keys will affect. Press Offset until the letter of the desired offset
is underlined.
Key Path: Meas Setup, Offset Setup
State Saved: Saved in instrument state.
Factory Preset: A
History: Added with PSA firmware revision A.02.00.
Remote Command:
There is no remote command for this function.
2.3.4.2 Offset Freq or Edge Freq
Enables you to set the frequency difference from the center of the main channel to the
center of the offset for a maximum of 6 offsets (labeled A-F). It also allows you to turn
on/off the offsets that you want to measure.
NOTE This key label changes to Edge Freq when IS-95C, cdma2000 MC-1X, cdma2000