Chapter 2 163
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
[:SENSe]:MCPower:CARRier:LIST:PPResent?
Remote Command Notes:
The position number in the list sent corresponds to the carrier number
you are designating. For example: “YES, YES, NO, YES, YES, NO”
defines six carriers. Carriers 1, 2, 4, and 5 are defined as having power
present whereas carriers 3 and 6 do not have power present. If you need
to change carrier 5, you must send all carriers up to 5. Carrier 6 will
remain unchanged. If you send more values than the number of carriers
specified using the Carriers key, (in this example, there are six carriers)
they will be ignored.
The query for this parameter returns the current value for all carriers
(with and without power present).
Example: MCP:CARR:PPR YES,NO,YES,NO,YES
MCP:CARR:PPR?
2.16.3.4.3 Carrier Width
Press Carrier Width to specify the width of each carrier including carriers with no power
present. There is a corresponding one-to-one relationship between each carrier width value
and the specified number of carriers. First, press the Carrier key and select the carrier
number you wish to define, using the Step Keys (⇓ ⇑), the knob, or the numeric keypad.
(The carrier number selected is shown on the Carrier key.) Then press the Carrier Width key
and enter the width using the numeric keypad.
Key Path: Meas Setup, Carrier Setup, Configure Carriers
Key Notes: The Carrier key determines which carrier width you are defining.
State Saved: Saved in instrument state.
Factory Preset: 5 MHz, 5 MHz, 5 MHz, 5 MHz
Terminators: Hz, kHz, MHz, GHz
Default Terminator: Hz
Resolution: 1 Hz
Step Key
Increment: If CF Step(Auto) is selected: span/10.
If CF Step(Man) is selected: CF Step
Range: 0 Hz to 500 MHz
History: Added with ESA firmware revision A.09.00.Added with PSA firmware
revision A.03.00.
Remote Command:
[:SENSe]:MCPower:CARRier:LIST:WIDTh <Hz>