168 Chapter 2
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
Added with ESA firmware revision A.08.00
Remote Command:
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA <dB>,<dB>,<dB>
:CALCulate:MCPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA?
Example: CALC:MCP:OFFS:LIST:LIM:NEG:DATA 50dB,55dB,60dB
CALC:MCP:OFFS:LIST:LIM:NEG:DATA?
2.16.5 Carrier Result
Press Carrier Result to select the result you wish to display on the last line of the carrier
power results list except when:
the carrier result number 4 (the first 4 carrier power results are displayed)
the carrier result number 9 (the last 4 carrier power results are displayed)
Key Path: Meas Setup
Key Notes: This key is only available when Meas Control, Measure (Single) is selected.
State Saved: Not saved.
Factory Preset: 1
Step Key
Increment: 1
Range: 1 to number of carriers.
History Added with PSA firmware revision A.03.00
Added with ESA firmware revision A.0900
Remote Command:
There is no equivalent remote command.

2.16.6 Optimize Ref Level

PSA Only: Sets the input attenuator to optimize the robustness of the measurement,
which is its freedom from errors due to input compression.
ESA Only: Sets the input attenuator and reference level to optimize the robustness of the
measurement, which is its freedom from errors due to input compression and log amp
range limitations.
NOTE There will always be a minimum of 2 dB for PSA or 5 dB for ESA of