154 Chapter 2
One-Button Measurement Functions
Meas Setup (Intermod (TOI))
One-Button Measurement Functions
[:SENSe]:TOIN:AVERage[:STATe]?
Example: TOIN:AVER ON
TOIN:AVER?
2.15.2 Avg Mode
Press Avg Mode to select the type of termination control used for the averaging function to
either Exp or Repeat. This determines the averaging action after the specified number of
measurements (average count) is reached.
EXP (Exponential Averaging mode)—When you set Avg Mode to Exp, each successive
data acquisition after the average count is reached is exponentially weighted and
combined with the existing average. Exponential averaging weights new data more
than old data, which facilitates tracking of slow-changing signals. The average will be
displayed at the end of each sweep.
Repeat—When you set Avg Mode to Repeat, after reaching the average count, all
previous result data is cleared and the average count is set back to 1.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: EXPonential
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00
Remote Command:
[:SENSe]:TOIN:AVERage:TCONrol EXPonential|REPeat
[:SENSe]:TOIN:AVERage:TCONrol?
Example: TOIN:AVG:TCON EXP
TOIN:AVG:TCON?
2.15.3 TOI Span
Press Span to specify the frequency span in which intermodulation products are measured.
Key Path: Meas Setup
Key Note: If you modify the value of Span in the base instrument Span menu, the
value in the Meas Setup menu will be updated to reflect the new value and
the measurement will restart if it is running.
Factory Preset: 15 MHz, or as defined by the radio standard selected.