156 Chapter 2
One-Button Measurement Functions
Meas Setup (Intermod (TOI))
One-Button Measurement Functions
TOIN:FREQ:MIX:RANG:AUTO 1
TOIN:FREQ:MIX:RANG:AUTO?
2.15.5 Optimize Ref Level
Sets the reference level to place the tones near the top of the display. When input
attenuation is set to Auto, pressing Optimize Ref Level sets the input attenuation according
to the auto rules described in Max Mixer Lvl, based on this reference level. You may want
to trade off analyzer noise and analyzer-induced distortion by manually changing the
attenuation setting from this starting point.
NOTE There is always a minimum of 2 dB for PSA or 5 dB for ESA of attenuation
set to protect the analyzer input.
Key Path: Meas Setup
State Saved: Saved in instrument state.
History: Added with PSA firmware revision A.02.00
Added with ESA firmware revision A.08.00
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example: TOIN:RANG:AUTO ONCE
TOIN:RANG:AUTO?