Chapter 2 93
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
values of these parameters if no radio standard (None) is selected.
Example: Sending fewer than six parameters to one of these commands will leave
the values of the unspecified offsets unchanged. If you don’t send settings
for all 6 offsets, it will set all the offsets that you specified, then it will set
any remaining offsets to the same setting as the last offset that you sent.
ACP:OFFS:LIST 50 Hz,75 Hz,100 Hz,125 Hz,150 Hz,175 Hz
ACP:OFFS:LIST:STAT ON,ON,ON,OFF,OFF,OFF
For example, after the above command is sent, sending the command with
only four paramenters (ACP:OFFS:LIST:STAT ON,ON,ON,OFF) will result
in the fifth and sixth offset remaining the same as the previous setting
(OFF).
2.3.4.3 Ref BW
Sets the reference bandwidth (integration bandwidth) for each offset.
Key Path: Meas Setup, Offset Setup
State Saved: Saved in instrument state.
Factory Preset: 2 MHz, if no radio standard (None) is selected.
If a radio standard is selected, the default value is defined by the selected
standard. This command, along with commands
[:SENSe]:ACPower:OFFSet:LIST[:FREQuency] and
[:SENSe]:ACPower:OFFSet:LIST:STATe are used to set an entire array of
values. The following table shows the default array if no radio standard
(None) is selected.
Offset State Frequency Integ. BW
A ON 3 MHz 2 MHz
BOFF0 2 MHz
COFF0 2 MHz
DOFF0 2 MHz
EOFF0 2 MHz
FOFF0 2 MHz
Offset State Frequency Integ. BW
A ON 3 MHz 2 MHz
B OFF 0 Hz 2 MHz