Reference, Installation, and Operations Manual

Section 3: Installation

3-9000-743 Rev S

June 2013

 

 

 

3.6.3Transducer assembly security seal installation

Use the following instructions to install the security seal wires on the Transducer Assembly (see Section 7.2 for transducer removal instructions).

Figure 3-29 Transducer housing, cable nut and security seal

A.

B.

C.

A.Transducer cable nutB.Transducer cable chordsetC.Security wire seal

1.Check the installation of the following parts:

Transducer Housing - securely tightened and flush with the meter body

Transducer cable nut - snugly tightened and flush with Transducer Housing

2.For each Transducer Assembly, install a security seal wire (maximum wire diameter 0.09 inch; 2.3mm) on the Transducer Cable Connector to the Transducer Cable Nut by rotating the Transducer Cable Nut clockwise; compressing the seal on the Transducer Cable Connector (see Figure 3-29).

3.Twist and adjust wire removing all slack and seal. Remove excess wire (see Figure 3-29).

Transducer assembly security seal installation