Reference, Installation, and Operations Manual

Section 6: Meter operation

3-9000-743 Rev S

June 2013

 

 

 

Table 6-18 Daily log content (Continued)

Data Point

Log action

 

 

PrevDayMacro1*:

Macro

bit 31

(unused)

bit 30

Is1BitMemoryError

bit 29

DI1

bits 28-26

(unused)

bit 25

IsDetectionErr

bit 24

IsCommErrAcqBd

bit 23

DidPowerFail

bit 22

DidCnfgChksumChg

bit 21

IsElecTempOutOfRange

bit 20

IsElecVoltOutOfRange

bit 19

IsGCAlarmPresent

bit 18

IsGCCommErr

bit 17

PressHARTIsCommErr

bit 16

TempHARTIsCommErr

bit 15

PressHartIsDevMalfunction

bit 14

TempHartIsDevMalfunction

bit 13

PressHARTIsConfigChanged

bit 12

TempHARTIsConfigChanged

bit 11

PressHARTDidColdStart

bit 10

TempHARTDidColdStart

bits 9-0

(unused)

 

 

PrevDayMacro2*:

Macro

bit 31

(unused)

bit 30

IsMeasSndSpdRange

bit 29

IsAvgSoundVelRangeErr

bit 28

IsMeterVelAboveMaxLmt

bit 27

IsHardFailedA

bit 26

IsHardFailedB

bit 25

IsHardFailedC

bit 24

IsHardFailedD

bit 23

(unused)

bit 22

IsSNRTooLow

bits 21-19

(unused)

bit 18

PressureInvalid

bit 17

TemperatureInvalid

bit 16

IsAnyLogFull

bit 15

(unused)

bit 14

IsGCWarningPresent

bit 13

PressHARTIsPVOutOfLimits

bit 12

TempHARTIsPVOutOfLimits

bit 11

PressHARTIsMoreStatusAvailable

bit 10

TempHARTIsMoreStatusAvailable

bit 9

IsBlockageDetected

bit 8

IsBoreBuildupDetected

bit 7

IsLiquidDetected

bit 6

IsAbnormalProfileDetected

bit 5

IsReverseFlowDetected

bit 4

IsSoundVelCompErr

bit 3-0

(unused)

 

 

Daily log

233