Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

2.Choose a Flow Calibration Method:

Polynomial

Piece-wise linear

None

3.Enter the appropriate forward and reverse meter factors and click the Write button to send the factors to the meter.

4.Click the Cancel button to close the dialog box. If you made changes, click the Write button to apply the changes.

After the meter is configured, set S-2 Position 4 (on the CPU Board) back to the open position to write-protect the configuration.

164

Set calibration parameters