Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

Table 5-15 Data Points for Analog (Current) Output Configuration

Daniel MeterLink

Data Points, Options and Guidelines

Display Name

 

 

 

 

Full scale mass rate:

Data points affected:

Series 100 Option Board

AO1FullScaleMassRate

Series 100 Plus Option Board

AO2FullScaleMassRate

(if installed)

Options:

 

Enter a mass rate (kg/hr or lbm/hr)

 

Guidelines

 

This point is only used when the analog output content is mass

 

 

rate.

 

This point specifies the mass rate associated with the maximum

 

 

current (20 mA).

 

For options and guidelines, see Section 6.2.3 and Daniel

 

 

MeterLink online help.

 

 

 

Alarm Action:

AO1ActionUponInvalidContent

Series 100 Option Board

AO2ActionUponInvalidContent

Series 100 Plus Option Board

Options:

(if installed)

High (0)

 

Low (1)

 

Hold Last Value (239)

 

Very Low - 3.5 mA (240)

 

Very High - 20.5 mA (241)

 

None (251)

 

 

 

Current Outputs Page