Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Table 5-11 Daniel MeterLink Field Setup Wizard

Page

Description

 

 

Frequency Outputs Page

The Frequency Outputs Page with the Series 100 Option Board allows you to configure the

Series 100 Option Board

frequency outputs. This page is only displayed if either the Frequency outputs was selected on the

See Section 5.6.7

Startup Page. Properties on this page are individually disabled according to selection made on the

 

Startup Page.

 

Content:

 

 

Uncorrected (Actual) flow rate

 

 

Corrected (Standard) flow rate

 

 

Average Flow Velocity

 

 

Average Sound Velocity

 

 

Energy Rate

 

 

Mass Rate

 

Direction:

 

 

These properties are only enabled if Frequency outputs or Current output was selected on

 

 

the Startup page. Select from the following options:

 

 

Reverse: The output only reports flow in the reverse direction. For frequency

 

 

 

outputs, Phase B of the output is 90 degrees out of phase with Phase A.

 

 

Forward: The output only reports flow in the forward direction. For frequency

 

 

 

outputs, Phase B of the output is 90 degrees out of phase with Phase A.

 

 

Absolute: The output reports flow in both directions. For frequency outputs, Phase B

 

 

 

of the output is 90 degrees out of phase with Phase A.

 

 

Bidirectional: The output reports flow on Phase A only in the forward direction and

 

 

 

on Phase B only in the reverse direction.

 

Channel B zero on error:

 

 

Select this check box if you want Phase B of the frequency output to go to zero while a

 

 

meter validity alarm is active. Phase A will continue to output as normal. If selected, Phase B

 

 

will go to zero when the associated output is put into Test mode indicating the meter

 

 

validity alarm is active.

 

Channel B phase:

 

 

Select the option to determine if you would either like Phase B to lag Phase A while

 

 

reporting forward flow and lead Phase A while reporting reverse flow or if you would like to

 

 

do the opposite. This selection only matters if the device to which the frequency outputs

 

 

are connected are configured for Dual Pulse security and phase detection. Else, selecting

 

 

either will have no effect on totalization.

 

Maximum frequency for frequency output:

 

 

Select the maximum frequency for the frequency outputs. Valid options are 1000 Hz and

 

 

5000 Hz. For Daniel Mark III Ultrasonic Gas Flow Meters, each frequency output has its own

 

 

register. This property is disabled if Frequency outputs was cleared on the Startup Page. The

 

 

Daniel Mark III Ultrasonic Gas Flow Meter will over range up to 7500 Hz.

 

 

 

 

126

Meter configuration using Daniel MeterLink