Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

1.To configure the Live Temperature, plus associated alarms, choose the settings listed in the table below):

Table 5-16 Live Temperature Inputs

Daniel MeterLink

Options and Guidelines

Display Name

 

 

 

Live temperature, Min input

Enter the temperature (°C or °F) that corresponds to a 4 mA input signal.

 

The temperature must be within the range [-273.15 °C, 200 °C].

 

 

Live temperature, Max input

Enter the temperature (°C or °F) that corresponds to a 20 mA input

 

signal. The temperature must be within the range [-273.15 °C, 200 °C].

 

 

Temperature alarm, Low limit

Enter a value (°C or °F) within the range [-130 °C, 200 °C]. An alarm is

 

generated when the temperature is at or below this limit value.

 

 

Temperature alarm, High limit

Enter a value (°C or °F) within the range [-130 °C, 200 °C]. An alarm is

 

generated when the temperature is at or above this limit value.

 

 

Alarm Selection

The Alarm selection determines what value to use while a live input is in

 

alarm. This value is common with the pressure Alarm selection so when

 

one is changed, the other will change to match.

 

Choose Last good value or Fixed value

 

 

150

Temperature and Pressure page