Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

7.Override the Mark III Serial Port B parameters to 19200 baud and Modbus ID 32 by setting CPU Board switch S2-1 to the CLOSED position (down toward the board) (or toggle the switch S2-1 if already in the CLOSED position).

Serial Port B remains in the override mode until either (1) the port configuration is changed (via software), (2) two minutes have elapsed at which time the port reverts to its current configuration), or (3) CPU Board switch S2-1 is changed back to the OPEN position.

8.From Daniel MeterLink Connect to Meter window, click the Direct button to connect to the meter record established above. Daniel MeterLink connects to the meter with the user-specified serial communication settings.

9.After connecting to the meter, configure the Mark III with the desired communication parameters via the Daniel MeterLink Meter>Communications Settings menu path.

10.Disconnect Daniel MeterLink from the Mark III meter.

11.Disable the Port B override by setting CPU Board switch S2-1 to the OPEN position (up away from the board).

At this point, the Daniel 3410 Series Ultrasonic Gas Flow Meter will only communicate at the user-configured settings.

12.In Daniel MeterLink, edit the meter directory record established above so that the Direct connection uses the user-specified serial communication settings. If desired, set the Modem and/or optional Ethernet connection settings to the user-specified values.

RS-485 serial connection setup

117