Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

5.6.11Meter Corrections page

Figure 5-14 Field Setup Wizard - Meter Corrections page

1.From the Field Setup Wizard - Meter Corrections Page, choose the desired settings:

a.Flow profile correction - Only enabled for Daniel JuniorSonic® Ultrasonic Gas Flow Meters and if Flow profile correction (for JuniorSonic® only) was selected on the Startup Page.

b.Flow profile correction factor is to be used - If Live or Fixed was selected for both temperature and pressure on the Startup Page, then you can choose from Calculated or Fixed for a correction factor. Selecting Fixed requires you to enter a value from 0.9 to 1.0 for Fixed correction factor. If Not used was selected for either temperature or pressure on the Startup Page, then you can choose from Fixed or Default for a correction factor. Calculated will be disabled. Selecting Fixed requires you to enter a value from 0.9 to 1.0 for Fixed correction factor. The default value is 0.95.

For Daniel Mark II Ultrasonic Gas Flow Meters only, the Fixed option is only available for DFI boards running version 3.72 or later firmware.

Daniel CUI version 4.21 is the final version that supports Daniel Mark II Gas Ultrasonic meters. If you are communicating with a Daniel Ultrasonic Mark II Gas Flow Meter, you do not need to uninstall version 4.21 or earlier versions of Daniel CUI before installing MeterLink. Version 4.21 and earlier versions of Daniel CUI should not be run simultaneously with MeterLink.

Meter Corrections page

147