Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Table 5-22 Data Points for Dry Calibration

Daniel MeterLink

Data Points, Options and Guidelines

Display Name

 

 

 

 

Forward A3

Data points affected:

 

FwdA3

 

Options:

 

Enter a value (s2/m2 or s2/ft2) within the range [-0.1,0.1 s2/m2].

 

Guidelines:

 

This value should only be modified under direction of Daniel personnel.

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

Reverse A0

Data points affected:

 

RevA0

 

Options:

 

Enter a value (m/s or ft/s) within the range [-1,1 m/s].

 

Guidelines:

 

This value should only be modified under direction of Daniel personnel.

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

Reverse A1

Data points affected:

 

RevA1

 

Options:

 

Enter a value (dimensionless) within the range [0.95,1.05].

 

Guidelines:

 

This value should only be modified under direction of Daniel personnel.

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

Reverse A2

Data points affected:

 

RevA2

 

Options:

 

Enter a value (s/m or s/ft) within the range [-0.1,0.1 s/m].

 

Guidelines:

 

This value should only be modified under direction of Daniel personnel.

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

Reverse A3

Data points affected:

 

RevA3

 

Options:

 

Enter a value (s2/m2 or s2/ft2) within the range [-0.1,0.1 s2/m2].

 

Guidelines:

 

This value should only be modified under direction of Daniel personnel.

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

 

166

Set calibration parameters