Section 6: Meter operation

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Normal meter operation continues while the new program is being downloaded and stored in FLASH memory. A new program generally does not affect the meter configuration and archive logs.

Once the program download is complete, the Mark III meter restarts in order to run the new program components. Daniel MeterLink issues a message that it must disconnect from the meter due to the meter re-start. Once the meter is re-started, re-connect to the meter with Daniel MeterLink.

6.8.1 Power disruption during firmware upgrade

If the program failed to download to a Daniel Mark III Ultrasonic Gas Flow Meter during a firmware upgrade, Daniel MeterLink can no longer connect to the meter or a message “Attempt FTP-only connection….” is displayed when trying to connect.

It is possible for the meter to lose its programming if the meter power fails during a firmware upgrade process. If this occurs, follow the special procedure known as a Backup Upgrade described below, to make a connection to the meter and re-attempt the download process.

1.In Daniel MeterLink, use the FileProgram Settings menu and select Allow FTP-only connection.

Figure 6-6 Program settings dialog FTP-only connection2.Click the OK button to close Program Settings dialog.

3.For Serial Port Connections: If you are connecting to the meter over a serial port, make sure you are connected to Port A or Port B on the meter Field Connection Board. Note:

You can not program the meter from Port C if the meter is not currently running a program.

256

Power disruption during firmware upgrade