Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

h.Use which gas composition on GC alarm: Select which gas composition the Daniel Mark III Gas Ultrasonic meter will use if the GC goes into alarm.

If Fixed value is selected, the meter will start using the fixed gas composition stored in the meter.

If Last good value is selected, the meter will use the last gas composition collected from the GC before the GC started to report alarms.

2.Click the Next button to continue the meter configuration. Daniel MeterLink displays the Field Setup Wizard - AGA-8 Page.

5.6.14AGA 8 Setup page

Figure 5-17 Field Setup Wizard - AGA8 Setup page

The Field Setup Wizard - AGA8 page is only displayed if both temperature and pressure are set to Live or Fixed and Base condition correction is selected in the earlier Field Setup Wizard - Startup page (see Section 5.6.5). Configuring AGA8 parameters is necessary for obtaining the desired AGA8 flow-condition compressibility calculation results and for optionally calculating the sound velocity (using AGA10 calculations).

AGA 8 Setup page

155