Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Temperature and Pressure Page Section 5.6.12Gas Chromatograph Setup Page Section 5.6.13AGA8 Setup Page Section 5.6.14Continuous Flow Analysis Page Section 5.6.15Update Time Wizard Section 5.7.1Table 5-11 Daniel MeterLink Field Setup Wizard

Page

Description

 

 

Startup

Meter name

See Section 5.6.5

Enter an alphanumeric name for the meter. When connecting, this name can be used as the

 

filename for logs and reports collected. This field is only available for Daniel Mark III Ultrasonic Gas

 

Flow Meters.

 

Serial #

 

Enter the serial number from the meter body. This will be included in the Inspection Report of

 

Maintenance logs. This field is only available for Daniel Mark III Ultrasonic Gas Flow Meters.

 

Station Name

 

Enter the station name for the meter’s location. This is included in the Inspection Report of Main-

 

tenance logs. This field is only available for Daniel Mark III Ultrasonic Gas Flow Meters.

 

Address

 

Enter the address for the meter’s location. This is included in the Inspection Report of Mainte-

 

nance logs. This field is only available for Daniel Mark III Ultrasonic Gas Flow Meters.

 

City

 

Enter the city for the meter’s location. This is included in the Inspection Report of Maintenance

 

logs. This field is only available for Daniel Mark III Ultrasonic Gas Flow Meters.

 

State and country

 

Enter the state and country for the meter’s location. This is included in the Inspection Report of

 

Maintenance logs. This field is only available for Daniel Mark III Ultrasonic Gas Flow Meters.

 

Temperature for meter corrections

 

Temperature is required for flow and for meter internal diameter Temperature expansion correc-

 

tion. If a temperature transmitter is connected to the meter, select Live for this option. If a temper-

 

ature transmitter is not connected to the meter but either Base condition correction or

 

Temperature expansion correction is required, select Fixed. You will enter in a fixed temperature

 

on a later page. If neither of the corrections is required, select Not used.

 

Pressure for meter corrections

 

Pressure is required for flow and for meter internal diameter Pressure expansion correction. If a

 

pressure transmitter is connected to the meter, select Live for this option. If a pressure transmitter

 

is not connected to the meter but either Base condition correction or Pressure expansion correc-

 

tion is required, select Fixed. You will enter in a fixed pressure on a later page. If neither of the cor-

 

rections is required, select Not used.

 

 

124

Meter configuration using Daniel MeterLink