Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

5.6.10Meter Digital outputs

Figure 5-13 Field Setup Wizard - Digital Outputs Page

1.From the Field Setup Wizard - Meter Digital Outputs Page, choose the desired settings:

a.Digital output group 1A, 1B, 2A, and 2BFor DO1AContent, choose Freq1 Validity (0) or Flow Direction (2)For DO1BContent, choose Freq1 Validity (0) or Flow Direction (2)For DO2AContent, choose Freq2 Validity (1) or Flow Direction (2)For DO2BContent, choose Freq2 Validity (1) or Flow Direction (2)For additional guidelines, see Section 5.6.7 and Daniel MeterLink online help.b.Inverted Operation (for any or all of the four digital outputs)

For DO1AContent, choose Freq1 Validity (0) or Flow Direction (2)

DO1AIsInvPolarityDO1BIsInvPolarityDO2AIsInvPolarity

DO2BIsInvPolarity unchecked - Normal (FALSE) checked - Inverted (TRUE) These data points allow the digital outputs polarity to be inverted from the normal (described in Section 5.6.7).

2.Click the Next button to continue the meter configuration. Daniel MeterLink displays the Field Setup Wizard - Meter Corrections Page.

146

Meter Digital outputs