Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

5.7.3Set calibration parameters

The Daniel Mark III Ultrasonic Gas Flow Meter meter provides for both “dry” calibration and “wet” calibration. Dry calibration refers to calibration methods that use factory-set calibration parameters. Wet calibration refers to calibration methods that use calibration parameters from a flow calibration.

Security switch S-2, Position 4 (on the CPU Board) must be closed or pushed towards the board before writing calibration factors to the meter.

1.Select the CalibrationMeter Factors menu and Daniel MeterLink displays the Meter Factors dialog.

Figure 5-21 Calibration - Meter Factors Page

Set calibration parameters