Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Table 5-25 Data Points for Polynomial Wet Calibration

Daniel MeterLink

Data Points, Options and Guidelines

Display Name

 

 

 

 

Reverse C1

Data points affected:

 

RevC1

 

Options:

 

Enter a value (dimensionless) within the range [0.95,1.05].

 

Guidelines:

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

Reverse C2

Data points affected:

 

RevC2

 

Options:

 

Enter a value (s/m or s/ft) within the range [-0.1,0.1 s/m].

 

Guidelines:

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

Reverse C3

Data points affected:

 

RevC3

 

Options:

 

Enter a value (s2/m2 or s2/ft2) within the range [-0.1,0.1 s2/m2].

 

Guidelines:

 

For additional guidelines, see Section 6.1.8 and Daniel MeterLink online help.

 

 

 

5.8Saving the meter configuration

It is recommended that, following meter set-up, the meter configuration be saved to a file for future reference. To save the configuration file,

1.Use the Tools}Edit/Compare Configuration menu.

2.Click the Meter Read button and, after the configuration data points are read, click the

File Save button.

170

Saving the meter configuration