Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

Pressure expansion correction: Only enabled if Live or Fixed was selected for Pressure for meter corrections and Pressure expansion correction was selected on the Startup Page.

Pipe outside diameter: Enter the outside diameter of the meter body.

Youngs modulus: Enter the Young's Modulus value (ratio of tensile stress to tensile strain)

Poissons ratio: Enter the Poisson's Ratio value (the absolute ratio of the pipe material lateral strain over axial strain)

 

For Daniel Mark II Gas Ultrasonic meters only, the Pressure expansion correction is only

 

available for DFI boards running version 3.10 or later firmware.

 

 

Table 5-12 Enabling Meter Corrections

 

 

 

 

Daniel MeterLink

Data Points, Options and Guidelines

Display Name

 

 

 

 

 

Base condition

Data points affected:

correction

None (enables/disabled access to other Daniel MeterLink configuration screens, data

 

 

 

fields)

 

 

Options:

 

 

Disabled if unchecked

 

 

Enabled if checked

 

 

Guidelines:

 

 

This selection enables/disables access to other Daniel MeterLink configuration screens

 

 

 

and/or data fields.

 

 

For options and guidelines, see Daniel MeterLink online help.

 

 

 

Flow profile correction

Data points affected:

(for JuniorSonicTM only)

None (enables/disables access to other Daniel MeterLink configuration screens, data

 

 

 

fields)

 

 

Options:

 

 

Disabled if unchecked

 

 

Enabled if checked

 

 

Guidelines:

 

 

For options and guidelines, see Daniel MeterLink online help.

 

 

 

Temperature expansion correc-

Data points affected:

tion

EnableExpCorrTemp

 

 

Options:

 

 

Disable (FALSE)

 

 

Enable (TRUE)

 

 

Guidelines:

 

 

For options and guidelines, see Daniel MeterLink online help.

 

 

 

 

Startup page

135