Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

c.Temperature expansion correction - Only enabled if Live or Fixed was selected for Temperature for meter corrections and Temperature expansion correction was selected on the Startup Page.

d.Linear expansion coefficient: Enter the linear expansion coefficient of the meter body material.

e.Linear expansion coefficient reference temperature: Enter the reference temperature for the Linear expansion coefficient.

For Daniel Mark II Ultrasonic Gas Flow Meters only, the Temperature expansion correction is only available for DFI boards running version 3.10 or later firmware.

See the Notice above regarding support for Mark II Gas Ultrasonic Flow Meters.

f.Pressure expansion correction - Only enabled if Live or Fixed was selected for Pressure for meter corrections and Pressure expansion correction was selected on the Startup Page.

g.Pipe outside diameter: Enter the outside diameter of the meter body.

h.Youngs modulus: Enter the Young's Modulus value (ratio of tensile stress to tensile strain)

i.Poissons ratio: Enter the Poisson's Ratio value (the absolute ratio of the pipe material lateral strain over axial strain)

For Daniel Mark II Ultrasonic Gas Flow Meters only, the Pressure expansion correction is only available for DFI boards running version 3.10 or later firmware.

2.Click the Next button to continue the meter configuration. Daniel MeterLink displays the Field Setup Wizard - Temperature and Pressure Page.

148

Meter Corrections page