Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Contract hour for daily log

Reverse flow: Select to Enable configure the meter to generate a reverse flow alarm if it accumulates more volume in the reverse direction than what is set in the Volume limit. For flow to be counted as reverse flow for this check, it must be flowing at a velocity about the Low flow limit. This low flow limit is different from the Volume cut-off threshold below which flow in not counted as flow for volume accumulation. Every time the flow direction changes from forward to reverse, the reverse flow accumulated volume is reset to zero. If the meter is a bi- directional meter, this alarm should not be enabled in order to avoid this alarm becoming active during periods of expected reverse flow.

3.Enter additional information as desired in the Notepad field.

4.Click the Next button to continue the meter configuration. Daniel MeterLink displays the Field Setup Wizard - Frequency Outputs Page.

5.6.7Frequency Outputs Page

Figure 5-10 Field Setup Wizard - Meter Frequency Outputs Page

138

Frequency Outputs Page