Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

Table 5-11 Daniel MeterLink Field Setup Wizard

Page

Description

 

 

 

Frequency Outputs Page

Volumetric flow

Series 100 Option Board

 

The following fields are used to configure the frequency or current outputs selected to

(continued)

 

output a volumetric flow rate. The fields are only enabled if the associated output’s Content

See Section 5.6.7

 

is set to Uncorrected (Actual) or Corrected (Standard).

 

Full scale volumetric flow rate used with output:

 

 

Enter the flow rate to be equivalent to the maximum frequency of the frequency output.

 

 

For Daniel Mark III Ultrasonic Gas Flow Meters, each frequency output has its own register.

 

 

This property is disabled if Frequency outputs was cleared on the Startup Page.

 

K-factor:

 

 

A read-only value showing the calculated K-factor from the Full scale volumetric flow rate

 

 

used with frequency outputs and the Maximum frequency for frequency output. This

 

 

property is disabled if Frequency outputs was cleared on the Startup Page.

 

Vol/pulse:

 

 

A read-only value showing the calculated inverse of the K-factor. This property is disabled if

 

 

Frequency outputs was cleared on the Startup Page.

 

Velocity

 

 

Maximum scale velocity used with output:

 

 

Enter the velocity to be equivalent to the maximum frequency of the frequency output.

 

 

These values are only enabled for Daniel Mark III Ultrasonic Gas Flow Meters and if the

 

 

Content is set to Average flow velocity or Average sound velocity. This property is disabled if

 

 

Frequency outputs was cleared on the Startup Page.

 

Minimum scale velocity used with output:

 

 

Enter the velocity to be equivalent to the minimum frequency (i.e. 0Hz) of the frequency

 

 

output. These values are only enabled for Daniel Mark III Ultrasonic Gas Flow Meters and if

 

 

the Content is set to Average flow velocity or Average sound velocity. This property is

 

 

disabled if Frequency outputs was cleared on the Startup Page.

 

 

 

Meter configuration using Daniel MeterLink

127