Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

Table 5-15 Data Points for Analog (Current) Output Configuration

Daniel MeterLink

Data Points, Options and Guidelines

Display Name

 

 

 

 

Content:

Data points affected:

Series 100 Option Board

AO1Content

Series 100 Plus Option Board

AO2Content

(if installed)

Options:

 

Uncorrected (Actual) flow rate (0)

 

Standard (corrected) flow rate (1)

 

Average flow velocity (2)

 

Average sound velocity (3)

 

Energy rate (4)

 

Mass rate (5)

 

Guidelines:

 

For options and guidelines, see Section 6.2.1 and Daniel

 

 

MeterLink online help.

 

 

Direction:

Data points affected:

Series 100 Option Board

AO1Dir

Series 100 Plus Option Board

AO2Dir

(if installed)

Options:

 

Reverse (0)

 

Forward (1)

 

Absolute (2)

 

Guidelines:

 

For options and guidelines, see Section 6.2.1 and Daniel

 

 

MeterLink online help.

 

 

Full scale volumetric flow rate:

Data points affected:

Series 100 Option Board

AO1FullScaleVolFlowRate

Series 100 Plus Option Board

AO2FullScaleVolFlowRate

(if installed)

Options:

 

Enter a flow rate value (m3/hour or ft3/hour)

 

Guidelines:

 

This point is only used when the current (analog) output content is

 

 

either the actual (uncorrected) or standard (corrected) flow rate.

 

This point specifies the volumetric flow rate associated with the

 

 

maximum current (20 mA).

 

For options and guidelines, see Section 6.2.1 and Daniel

 

 

MeterLink online help.

 

 

 

Current Outputs Page

141