Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Table 5-21 Data Points for Batch Cycle Processing, or Update Time Control

Daniel MeterLink

Data Points, Options and Guidelines

Display Name

 

 

 

 

Smoothing%

Data points affected:

 

BatchPercentSmoothing

 

Options:

 

0% (0)

 

20% (20)

 

40% (40)

 

60% (60)

 

80% (80)

 

Guidelines:

 

For additional guidelines, see Section 6.1.3 and Daniel MeterLink online

 

 

help.

 

 

 

5.7.2Calculated results

The Daniel MeterLink Update Time Wizard displays the following meter-calculated values:

Update time

Samples/update

Emission rate

Stacked Emission rate

Past samples used

Total samples used

The samples per update is the average number of new samples (waveforms per transducer) collected during an update period. The past samples used shows how many samples from previous update periods will be used in the current update period. The total samples used is the total of past and new samples used in the current update period. Refer to Section 6.1.3 for further information on these update-time related values.

Beginning with Daniel MeterLink version 3.10, the Update Time Wizard also displays the path signal-to-noise ratios (SNRs) and waveforms to facilitate determining if applying stacking and/or filtering improves the signals. Generally, signal improvement is indicated by an increase in the SNR. Note that the SNRs and waveforms are for the parameters currently used by the meter; parameter changes do not take affect until the Write button is clicked.

162

Calculated results