Section 6: Meter operation

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Profile Factor

The profile factor is a dimensionless ratio of the inner chord velocities to the outer chord velocities. It is readable via the ProfileFactor data point. Ideally the Profile Factor should be 1.0 for Dual-X meters (model 3422) and 1.17 for British Gas-style meters (model 3400).

Swirl angle

The swirl angle is an estimate of the swirl (to the nearest degree) and is readable via the SwirlAngle data point. It is calculated as a function of the meter body style and Profile Factor (see above). A value of 0 degrees indicates no appreciable swirl.

6.1.15Flow-condition pressure and temperature

The flow-condition pressure and temperature are used by the Daniel Mark III Ultrasonic Gas Flow Meter for various calculations such as

expansion correctionflow profile correction (JuniorSonicmeters only)calculation of base (standard) condition volumetric flow rate and volumesoptional AGA10 sound velocity calculationConfiguration

The flow-condition pressure and temperature are individually configurable (via the EnablePres- sureInput and EnableTemperatureInput data points) to be

disabled (0),live (1) (4-20 mA input signal, requires the Option Board), orfixed (2).

If an input is live, then the values corresponding to the minimum and maximum input (4 and 20 mA, respectively) are specified via data points (MinInputPressure, MaxInputPressure, MinInputTemperature, MaxInputTemperature).

If an input is fixed, then its value is specified via a data point (SpecFlowPressure, SpecFlow- Temperature).

Alarm limits can be specified for each input (LowPressureAlarm, HighPressureAlarm, Low- TemperatureAlarm, HighTemperatureAlarm). Additionally, the flow-condition pressure is configurable to be gage or absolute (via the InputPressureUnit data point). If the pressure is gage, then the atmospheric pressure must be specified (via the AtmosphericPress data point). See Section 5.6.12 for details on configuring the flow-condition pressure and temperature.

Another data point (FlowPOrTSrcUponAlarm), common to both pressure and temperature, is used to specify the data source to use when the selected input data is invalid (i.e., value at or outside its alarm limits or a live input in calibration) as either

last good value (0), orfixed value (1).

This data point (FlowPOrTSrcUponAlarm) is not configurable via the Daniel MeterLink Field Setup Wizard; however, it is configurable via the Daniel MeterLink Edit/Compare Configuration screen (ToolsEdit/Compare Configuration). The default is to use the last good value.

206

Flow-condition pressure and temperature