Section 6: Meter operation

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

 

Digital output test mode

 

The Mark III meter provides a mode of operation for individually testing the digital output signal

 

pairs. Entering, configuring, and exiting this mode is accomplished by setting data points using

 

the Daniel MeterLink Tools - Outputs Test screen as described in the table below.

Table 6-13 Data points for digital outputs test

 

 

 

 

Daniel MeterLink

Data Points, Options and Guidelines

 

Display Name

 

 

 

 

 

 

 

Digital output 1

Data points affected:

 

Output 1A

DO1ATestVal

 

or

DO1BTestVal

 

Digital output 1

DO2ATestVal

 

Output 1B

DO2BTestVal

 

or

 

Options:

 

Digital output 2

 

Select either Test Low(0) to force the corresponding output low or Test High(1) to

 

Output 2A

 

 

force the corresponding output high.

 

or

 

 

Guidelines:

 

Digital output 2

 

The selected test value takes effect within one batch period of clicking the

 

Output 2B

 

 

Start DO1 or Start DO2 button (see below).

 

 

 

 

 

For options and guidelines, see timeout discussion below and Daniel MeterLink

 

 

 

online help.

 

 

 

 

Start(Stop) DO1

Data points affected:

 

or

DO1PairTestEnable

 

Start(Stop) DO2

DO2PairTestEnable

 

 

Options:

 

 

Click on the Start button to enter the test mode (TRUE) (Note: the Start button will

 

 

 

become the Stop button after clicking on it.)

 

 

Click on the Stop button to exit the test mode (FALSE) (Note: the Stop button will

 

 

 

become the Start button after clicking on it.)

 

 

Guidelines:

 

 

The selected test value (see above) takes effect within one batch period of clicking

 

 

 

the Start DO1 or Start DO2 button.

 

 

For further information, see timeout discussion below and Daniel MeterLink online

 

 

 

help.

 

 

 

 

 

The maximum length of time that a digital output pair can remain in the test mode is specified

 

via the NonNormalModeTimeout data point. Note that this data point applies to both digital

 

output pairs (as well as various other tests). The NonNormalModeTimeout data point can be

 

changed using the Daniel MeterLink Edit/Compare Configuration screen. It can be set within the

 

range [1, 30 min] and has a default value of 2 min.

220

Digital outputs