Reference, Installation, and Operations Manual

Section 5: Startup and meter configuration

3-9000-743 Rev S

June 2013

 

 

 

2.If Fixed was selected for Temperature for meter corrections on the Startup Page, enter an average temperature of the process fluid:

Table 5-17 Fixed Temperature Inputs

Daniel MeterLink

Options and Guidelines

Display Name

 

 

 

Fixed temperature

Enter the flow-condition temperature (°C or °F) within the range

 

[-130 °C, 200 °C]

 

 

Temperature alarm, Low limit

Enter a value (°C or °F) within the range [-130 °C, 200 °C]. An alarm is

 

generated when the temperature is at or below this limit value.

 

 

Temperature alarm, High limit

Enter a value (°C or °F) within the range [-130 °C, 200 °C]. An alarm is

 

generated when the temperature is at or above this limit value.

 

 

Alarm Selection

The Alarm selection determines what value to use while a live input is

 

in alarm. This value is common with the pressure Alarm selection so

 

when one is changed, the other will change to match.

 

Choose Last good value or Fixed value

 

 

Temperature and Pressure page