Section 5: Startup and meter configuration

Reference, Installation, and Operations Manual

June 2013

3-9000-743 Rev S

 

 

Ethernet Initial Connection Material Checklist

To complete the steps in the subsections that follow, obtain the following materials and information:

Daniel 3410 Series Gas Ultrasonic Flow Meter Ethernet adapter cable (Daniel P/N 3-3400-079)

Personal computer (PC) configured as follows:Daniel MeterLink software installed (version 3.0 or higher)Ethernet LAN adapterConfigure to automatically obtain an IP address (via DHCP)Desired meter(s) communication configuration parameters:IP addressserial communication parameters such as baud rate and Modbus ID (if desired)

5.3.4Ethernet initial connection steps

1.Power up the meter.2.Shutdown the PC.

3.Plug the Ethernet adapter cable Phoenix end into the meter Field Connection Board connector J8 and connect the RJ-45 end into the PC Ethernet connector.

4.Enable the Ethernet LAN connector DHCP server on the CPU Module by moving the DHCP (switch-1) to the ON position (see direction arrow on the CPU Module label).

5.Power up (boot) the PC and log in to the initial Windows logon prompt.

6.Verify the Ethernet connection status by the CPU Module “LINK” LED which should be on solid green.

7.Launch Daniel MeterLink.