
Section 4 System Start Up
Important Note: During initial commissioning, make sure all
1.Supply power to the controller. When the LED light turns green, the display module and the attached devices are communicating.
2.Follow the touch screen calibration prompts. After the touch screen calibration is complete the operating system starts and the display automatically prompts for the user language, time and date.
Note: Touch screen calibration is required for each user. Calibrating the system to a stylus will prevent the need for multiple operator calibrations. The initial touch screen calibration is stored in the display module. To change the touch screen calibration, switch the display module on and off. Press the screen during
3.Select the appropriate language and time and date settings.
4.Switch the display module off and on.
5.Confirm the attached probes and devices.
6.Press OK.
7.The controller automatically scans for connected probes. The scan may take several minutes to complete.
For more information on using the display module, refer to section 5.1 on page 53.
51