Advanced Operations
6.3.3.4 Function set to 2 POINT CONTROL working mode
2 POINT CONTROL
|
|
| Default value: |
| SET TRANSFER | Sets the status of the relay | |
|
|
| the selected source or if the source is missing. |
|
|
|
|
|
|
| Default value: HIGH |
| PHASE | Sets the status of the relay. As soon as the process value enters the band between high and low | |
|
|
| alarm, the relay state is not changed. |
|
|
|
|
|
| HIGH | Turns the relay ON when the process value exceeds the high alarm |
|
| Turns the relay OFF when the process value falls below the low alarm. | |
|
|
| |
|
|
|
|
|
| LOW | Turns the relay ON when the process value falls below the low alarm. |
|
| Turns the relay OFF when the process value exceeds the high alarm. | |
|
|
| |
|
|
|
|
| HIGH ALARM | Default value: 15 | |
| Sets the upper limit in the unit of the selected parameter of the 2 point control band. | ||
|
|
| |
|
|
|
|
| LOW ALARM | Default value: 5 | |
| Sets the lower limit in the unit of the selected parameter of the 2 point control band. | ||
|
|
| |
|
|
| |
| ON DELAY | Default value: 5 seconds | |
| (0 | Sets a delay time for the relay to turn on. | |
|
|
| |
| OFF DELAY | Default value: 5 seconds | |
| (0 | Sets a delay time for the relay to turn off. | |
|
|
|
|
|
|
| Default value: 0 minutes (off) |
| OnMax TIMER | Sets a maximum period of time. During this period the relay is turned ON when passing the | |
| corresponding limit. As soon as the time expires the relay is turned OFF regardless of the | ||
| (0 | ||
| process value. | ||
|
|
| |
|
|
| 0=OnMax Timer is not active. |
|
|
|
|
|
|
| Default value: 0 minutes (off) |
| OffMax TIMER | Sets a maximum period of time (in minutes). During this period the relay is turned OFF when | |
| passing the corresponding limit. As soon as the time expires the relay is turned ON regardless of | ||
| (0 | ||
| the process value. | ||
|
|
| |
|
|
| 0=OffMax Timer is not active. |
|
|
|
|
|
|
| Default value: 0 minutes (off) |
| OnMin TIMER | Sets a minimum period of time. During this period the relay is turned ON when passing the | |
| corresponding limit. The relay can only be turned OFF after the period has expired and after this | ||
| (0 | ||
| expiration will be turned OFF dependent on the process value. | ||
|
|
| |
|
|
| 0=OnMin Timer is not active. |
|
|
|
|
|
|
| Default value: 0 minutes (off) |
| OffMin TIMER | Sets a minimum period of time. During this period the relay is turned OFF when passing the | |
| corresponding limit. The relay can only be turned ON after the period has expired and after this | ||
| (0 | ||
| expiration will be turned ON dependent on the process value. | ||
|
|
| |
|
|
| 0=OffMin Timer is not active. |
|
|
|
|
|
|
| Default value: 0 seconds (off) |
|
|
| Indicates a period of time (in seconds) for the expiration of the OnMax TIMER and the OffMax |
|
|
| TIMER. |
| MAX TIMER EXPIRE | Relay turned on, OnMax TIMER activated: The time left is displayed before the relay is turned off | |
|
|
| automatically. |
|
|
| Relay turned off, OffMax TIMER activated: The time left is displayed before the relay is turned on |
|
|
| again. |
|
|
|
|