Installation
3.4 ![](/images/new-backgrounds/1192690/19269035xi2.webp)
Electrical installation
DANGER
Electrocution hazard. Only qualified personnel should conduct the installation tasks described in this section of the manual.
DANGER
Electrocution hazard. Always install a ground fault interrupt circuit (GFIC)/ residual current circuit breaker (rccb) with a maximum trigger current of 30 mA. If installed outside, provide overvoltage protection.
If installed outdoors, provide overvoltage protection between the power and the sc1000 controller. Make sure the data and power cables do not cause tripping and do not contain any sharp bends. Refer to Figure 7 for housing breakout information.
The instrument can be wired for line power by
Do not connect the electrical power supply to the AC power until the sc1000 controller has been fully wired, fused, and the high voltage barrier and probe module cover have been replaced.
3.4.1 Installation in hard-wired applications
In
Note: There is no on/off switch to disconnect the probe module from AC Power.