6.3.4 Network Modules (Profibus, Modbus)

The sc1000 controller can be embedded as a slave into an existing fieldbus system. The Network Modules menu displays all the settings needed, the menu content depends on the communication gateway used, Profibus DP or Modbus.

Note: This menu is displayed only if a network card is installed in the sc1000 controller.

6.3.4.1 Profibus

SYSTEM SETUP

NETWORK MODULES

FIELDBUS

 

 

 

Builds an individual composition of data from different devices. These data structure allows to

 

TELEGRAM

transmit up to 24 measured values on one Profibus slave device.

 

 

 

For details refer to section 5.12 on page 62.

 

 

 

 

PROFIBUS DP

 

 

 

 

 

 

 

ADDRESS

Default value: 0

 

 

Sets up the PROFIBUS address (1 to 128) for the slave.

 

 

 

 

 

 

 

 

 

 

Default value: NORMAL

 

 

 

Sets the sequence of bytes when transmitting floating point values.

 

 

 

Please note that this setting only affects the data of the configured slave.

 

 

DATA ORDER

A floating point value consists of 4 bytes.

 

 

SWAPPED: Swaps the first pair of bytes with the last pair.

 

 

 

 

 

 

NORMAL: The pairs are not swapped. This mode fits to all known Profibus master systems.

 

 

 

Note: A wrong setting in this menu can lead to slight deviations of floating point values shifted by

 

 

 

one register.

 

 

 

 

 

 

 

Simulates two floating point values and error/status to substitute a real instrument.

 

 

 

The tag order is:

 

 

 

1. Tag: ERROR

 

SIMULATION

2. Tag: STATUS

 

3./4. Tag: First floating point value counting to MAXIMUM value respectively MINIMUM value.

 

 

 

5./6. Tag: Second floating point being the difference between the first floating point value and the

 

 

 

value set in the MAXIMUM. menu

 

 

 

The first floating point runs through a ramp between the limits set in the menus MAXIMUM and

 

 

 

MINIMUM. Figure 65 shows the simulation mode.

 

 

 

 

 

 

 

Default value: NO

 

 

SIMULATION

Turns the simulation on or off.

 

 

YES: Starts a simulation

 

 

 

 

 

 

NO: Stops a simulation.

 

 

 

 

 

 

 

Default value: 10 minutes

 

 

PERIOD

Sets the time the first floating point value needs to run through the whole range between

 

 

 

MINIMUM and MAXIMUM.

 

 

 

 

 

 

MAXIMUM

Default value: 100

 

 

Sets the upper limit for the first floating point value.

 

 

 

 

 

 

 

 

 

MINIMUM

Default value: 50

 

 

Sets the lower limit for the first floating point value.

 

 

 

 

 

 

 

 

 

ERROR

Default value: 0

 

 

The value entered in this menu will be set in the first simulated tag (Table 15).

 

 

 

 

 

 

 

 

 

STATUS

Default value: 0

 

 

The value entered in this menu will be set in the second simulated tag (Table 16).

 

 

 

 

 

 

 

 

 

TOGGLE

Changes the direction of the simulated ramp.

 

 

 

 

109

Page 111
Image 111