Hach sc1000 user manual Function set to Alarm working mode

Models: sc1000

1 150
Download 150 pages 30.78 Kb
Page 90
Image 90

Advanced Operations

SYSTEM SETUP

RELAY

RELAY INT/EXT

 

 

FEEDER CONTROL

Operates in response to the measured parameter. Can be set for phasing, set point, deadband,

 

 

 

overfeed timer, and ON/OFF delay.

 

 

2 POINT CONTROL

Operates in response to the measured parameter using two setpoints.

 

 

 

 

 

 

WARNING

Activated when the analyzer detects a probe warning. Indicates warning and error condition of

 

 

selected probes.

 

 

 

 

 

 

 

 

 

PWM Control

Allows the relay to provide a pulse width modulated output.

 

 

 

 

 

 

FREQ. CONTROL

Allows the relay to cycle at a frequency between the minimum pulses per minute and maximum

 

 

pulses per minute.

 

 

 

 

 

TIMER

Allows the relay to switch at certain times independent of any process value

 

 

 

 

 

 

SYSTEM ERROR

Indicates if a probe in the system has an internal error or warning

 

 

 

 

 

INPUT VALUE

The process value read from the selected source after it has been processed by the internal

 

formula parser (if prompted).

 

 

 

 

 

 

 

 

 

 

Default value: OFF

 

LOG INTERVAL

Sets the interval for logging the displayed value into the data logger.

 

 

 

Options: OFF, 5min, 10min, 15min, 20min, 30min

 

 

 

 

6.3.3.2 Function set to ALARM working mode

ALARM

 

 

 

Default value: DE-ENERGIZED

 

SET TRANSFER

Sets the status of the relay (ENERGIZED/DE-ENERGIZED) if error conditions are detected in

 

 

 

the selected source or if the source is missing.

 

 

 

 

 

PHASE

Default value: DIRECT

 

Decides if the relay is turned on or off when the process value is leaving the controlled band.

 

 

 

 

 

 

 

 

 

DIRECT

Relay is turned on when leaving the controlled band

 

 

 

 

 

 

REVERSE

Relay is turned off when leaving the controlled band

 

 

 

 

 

HIGH ALARM

Default value: 15

 

Sets the maximum of the controlled band in the selected parameter unit.

 

 

 

 

 

 

 

 

LOW ALARM

Default value: 5

 

Sets the minimum of the controlled band in the selected parameter unit.

 

 

 

 

 

 

 

 

HIGH DEADBAND

Default value: 1

 

Sets the hysteresis value used at the upper limit.

 

 

 

 

 

 

 

 

LOW DEADBAND

Default value: 1

 

Sets the hysteresis value used at the lower limit

 

 

 

 

 

 

 

ON DELAY

Default value: 5 seconds

 

(0 s–999 s)

Sets a delay time for the relay to turn on.

 

 

 

 

OFF DELAY

Default value: 5 seconds

 

(0 s–999 s)

Sets a delay time for the relay to turn off.

 

 

 

 

88

Page 90
Image 90
Hach sc1000 user manual Function set to Alarm working mode