Advanced Operations
SYSTEM SETUP
RELAY
RELAY INT/EXT
|
| FEEDER CONTROL | Operates in response to the measured parameter. Can be set for phasing, set point, deadband, |
|
|
| overfeed timer, and ON/OFF delay. |
|
| 2 POINT CONTROL | Operates in response to the measured parameter using two setpoints. |
|
|
|
|
|
| WARNING | Activated when the analyzer detects a probe warning. Indicates warning and error condition of |
|
| selected probes. | |
|
|
| |
|
|
|
|
|
| PWM Control | Allows the relay to provide a pulse width modulated output. |
|
|
|
|
|
| FREQ. CONTROL | Allows the relay to cycle at a frequency between the minimum pulses per minute and maximum |
|
| pulses per minute. | |
|
|
| |
|
| TIMER | Allows the relay to switch at certain times independent of any process value |
|
|
|
|
|
| SYSTEM ERROR | Indicates if a probe in the system has an internal error or warning |
|
|
|
|
| INPUT VALUE | The process value read from the selected source after it has been processed by the internal | |
| formula parser (if prompted). | ||
|
|
| |
|
|
|
|
|
|
| Default value: OFF |
| LOG INTERVAL | Sets the interval for logging the displayed value into the data logger. | |
|
|
| Options: OFF, 5min, 10min, 15min, 20min, 30min |
|
|
|
|
6.3.3.2 Function set to ALARM working mode
ALARM
|
|
| Default value: |
| SET TRANSFER | Sets the status of the relay | |
|
|
| the selected source or if the source is missing. |
|
|
|
|
| PHASE | Default value: DIRECT | |
| Decides if the relay is turned on or off when the process value is leaving the controlled band. | ||
|
|
| |
|
|
|
|
|
| DIRECT | Relay is turned on when leaving the controlled band |
|
|
|
|
|
| REVERSE | Relay is turned off when leaving the controlled band |
|
|
|
|
| HIGH ALARM | Default value: 15 | |
| Sets the maximum of the controlled band in the selected parameter unit. | ||
|
|
| |
|
|
|
|
| LOW ALARM | Default value: 5 | |
| Sets the minimum of the controlled band in the selected parameter unit. | ||
|
|
| |
|
|
|
|
| HIGH DEADBAND | Default value: 1 | |
| Sets the hysteresis value used at the upper limit. | ||
|
|
| |
|
|
|
|
| LOW DEADBAND | Default value: 1 | |
| Sets the hysteresis value used at the lower limit | ||
|
|
| |
|
|
| |
| ON DELAY | Default value: 5 seconds | |
| (0 | Sets a delay time for the relay to turn on. | |
|
|
| |
| OFF DELAY | Default value: 5 seconds | |
| (0 | Sets a delay time for the relay to turn off. | |
|
|
|
|
88